| Specification \ Model |
H3-60 |
Sx60 |
Hx40 |
Hx60 |
| Available |
now |
now |
now |
now |
FOM, spherical section
(radius x width x height) |
240 x 200 x 160 cm |
115 x 70 x 55 cm |
120 x 120 x 90 cm |
200 x 130 x 100 cm |
| Measurement rate |
16 Hz |
48 Hz |
20 Hz |
20 Hz |
| Calibration accuracy1 |
0.20 mm RMS |
0.25 mm RMS |
0.20 mm RMS |
0.35 mm RMS |
| Jitter, static target2 |
0.007 mm RMS |
0.007 mm RMS |
0.015 mm RMS |
0.015 mm RMS |
| Jitter, moving target2 |
0.07 mm RMS |
0.07 mm RMS |
0.14 mm RMS |
0.14 mm RMS |
Processing time/frame
(2.5GHz Intel Q8300, 4 markers) |
15 ms |
7 ms |
10 ms |
10 ms |
| Lag |
~60 ms |
~20 ms |
~50 ms |
~50 ms |
| Sensors resolution |
1280x960 |
640x480 |
1024x768 |
1024x768 |
| Lenses / H°xV° |
6 mm 50°x38° |
6 mm 50°x38° |
4 mm 70°x52° |
6 mm 50°x38° |
| Case dimensions (WxHxD, approx.) |
283 x 43 x 49 mm |
164 x 43 x 54 mm |
164 x 43 x 54 mm |
164 x 43 x 54 mm |
| Weight (approx.) |
505g |
342g |
342g |
342g |
| Electrical Interface |
IEEE-1394b (FireWire), 800 Mbps |
IEEE-1394a (FireWire), 400 Mbps |
| Mount |
1/4" thread tripod mount |
| Operating temperature |
15-35°C (59-95°F) |
| Warm-up period |
~20 minutes |
~15 minutes |
| Max. # of marker templates |
practically unlimited |
| Max. # of markers concurrently tracked |
100 |
| Illumination range |
50-100,000 Lux (20-400,000 Lux in HDR mode) |
| Max illumination variation between markers |
6:13 (1000:1 in HDR mode) |
| Certifications |
Class II medical device certifications: IEC 60601-1, CE and FCC Part 15 Class A, RoHS and WEEE compliant |
| Drift: temperature, pressure, light intensity, marker orientation, motion blur, exposure settings |
Insignificant |
| Hazard warnings issued |
Marker out of FOM; Thermal instability; Temperature out of range; Shadow over target |
| Operating system |
MS-Windows 2000/XP/Vista/Windows 7, Linux, Mac OSX - 32bit and 64bit |
| Minimal computer hardware |
2GHz CPU, 512 MB RAM, 500MB disk space |
| |
|
| |
|
1. Hx40, single target (Xpoint), over 20,000 averaged positions at depths of 40-100cm
2. Hx40, single target at a depth of 75cm, 150 Lux, 20ms shutter
3. The range can be extended to 20:1 with black/gray targets under bright illumination
|
| |
|
 |
| |